TS SERIES

These sterilizers come equipped with a built-in ventilation system that combines clean steam with air to produce the sterilizing agent. This technique is advised for intricate plastic or glass containers like pre-filled syringes or vials, where condensation might accumulate and potentially affect the end product. The system guarantees consistent heat distribution during the entire sterilization process, with a temperature variation of less than ± 1°C. Additionally, these autoclaves are highly efficient

  • The TS-AS sterilization chamber features a distinctive circular shape with a flat cavity, facilitating fast and uniform heating/cooling processes, and is equipped with square or rectangular doors for accessibility.
  • The TS-OW model is constructed with a single-wall design, lacking the aforementioned cavity.
  • Stability during use is ensured through reinforcements welded to the exterior of both the chamber and cavity.
  • A custom-designed base supports the chamber, promoting even weight distribution across the floor.
  • The chamber incorporates height-adjustable guides at its base for the insertion and removal of loading trolleys.
  • Efficient condensate disposal is aided by the chamber's bottom design, which slopes towards a drain sump.
  • Two 2-inch tri-clamp ferrules are attached to the chamber, serving validation needs like temperature uniformity mapping and pressure tests.
  • All internal surfaces of the chamber undergo mechanical polishing, chemical degreasing, pickling, and passivation treatments, achieving a surface finish of ≤ 0.35 microns.
  • Depending on the chamber's size, installation can be either above-floor or recessed.

Products

Heat sensitive materials such as plastic syringes, infusion sets, dialysis filter cartridges, plastic materials and special surgical tools.

Capacity

 

Door type

Sliding/Hinged

 